Maintenance

9.1

Maintenance schedule

Part
Filters
Oxygen sensor
Oxygen sensor
Table 9-1: Maintenance schedule
9.2

Calibrate oxygen sensor

1. Remove the sensor cap (E) and expose the sensor to ambient air.
2. Wait for 60 seconds.
3. Then enter again the maintenance menu
3/5 wait until the oxygen level is stable.
4. Then go to screen 5/5 and touch O2 20.9% button.
The system will ask for confirmation. Select YES.
5. Reconnect the tube (F) and the sensor cap (E) with the sensor (C).
9.3

Cleaning

Clean the equipment with a damp cloth only and avoid excessive moisture around any electrical sockets. If
required you may use a mild detergent, however do not use abrasives or solvents as they may damage the
warning labels on the equipment.
K3.1.142M Manual NitroFlow Lab
Action
Replace carbon adsorber
Replace oxygen sensor
Calibrate oxygen sensor
Frequency
, screen
Fig. 9-8: Calibrate oxygen sensor
1x per year
1x per 3 years
1x 3 months
- 29 -