Options Menu; Local Settings Menu - Parker NitroFlow Lab Guia Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

6.3.4

Options menu

Access:
Press option menu button
Function:
Set different options
ATTENTION:
All options are default set to NO. Options do not affect the output and purity.
Button
Auto R.
Auto R.
Remote
Remote
Pincode
Pin code
Pin code
Show Flow
Show Flow (D)
6.3.5

Local settings menu

Access:
Touch local settings menu button
(refer to § 6.3)
Function:
Set data to local requirements
Depending on the local situation it is possible to change the setting accordingly.
K3.1.142M Manual NitroFlow Lab
in settings screen (refer to §6.3)
Selection
Result
Yes
After a power failure the unit will automatically restart itself and
return to the same situation/status.
No
After a power failure the unit will not start automatically. Unit
needs to be restarted manually.
Yes
Unit can be switch on and off from a remote location. Only
select Yes after connecting the printed circuit board to an
external device.
No
Unit cannot be controlled from a remote location.
Yes
Settings are instantly protected with a pin code. Return to log
on menu
No
Settings can be changed without a pin code
Pin code can be changed to a personal 4 digits code. (In case
Change
you forget your personal code, consult your supplier)
Yes
Flow rate will be displayed in main screen
No
Flow rate will not be displayed.
Operate to adjust date and time
and enter the default pin code 1234.
in settings screen
- 21 -