6 Preparation for use
WARNINGS
For use only by medically trained persons who have been adequately trained in suction
procedures and in the use of aspirators�
CAUTIONS
– The Dominant Flex must remain in an upright position during use�
– The rack version requires a minimum distance of 5 cm to the enclosure to prevent from
overheating of the device� The back of the enclosure must be open�
– Sterile accessories must be checked to ensure the integrity of the packaging before use�
– Non sterile and reusable accessories must be cleaned, disinfected and / or sterilized
according to the Medela cleaning guide (product code 200�2391)�
6.1
Checks before use
6.2
Assembly of the
basic configuration
– Check the Dominant Flex system before use for damage
of the power cord or plug, obvious device damage or
safety defects and proper functioning of the device.
– Check for completeness and general condition of the
Dominant Flex delivery package.
– Check all accessories prior to use:
– suction jars, lids and liners for cracks, brittle and
flawed spots. Replace if necessary.
– Tubing for cracks, brittle areas and that connectors
are firmly attached. Replace if necessary.
– As an additional safety test, evacuate the system
(including jars) to maximum vacuum before actual use.
1�
2�
3� Attach all necessary accessories according to your needs�
See chapters 17 / 18�
1�1 Make sure that the Safety
Set is attached to the
Dominant Flex�
See chapter 5�5�
2�1 Attach a filter (if required)
to the Safety Set with the
arrow pointing in the flow
direction� The filter must
be exchanged after an
overflow and after each
patient, but latest after a
week�