Preparation for use
WARNINGS
For use only by persons who have been adequately
trained in the use of suction pumps for airway
suctioning by their healthcare provider / physician.
CAUTIONS
– The Vario 18 must remain in an upright position
during use.
– Sterile accessories must be checked on the
integrity of the packaging before use.
– Non sterile and reusable accessories must be
cleaned and disinfected before use.
Checks before use
– Check the Vario 18 pump before use for damage
of the power cord or plug, obvious device damage
or safety defects and proper functioning of the
device.
– Only for AC / DC versions: make sure that the
rechargeable battery is calibrated according to
chapter «battery calibration».
– Check all accessories prior to use:
– suction jars, lids and liners for cracks, brittle
and flawed spots. Replace if necessary.
– Tubing for cracks, brittle areas and that connec-
tors are firmly attached. Replace if necessary.
11