Connection A - High Vacuum Flange; Connection Configurations - Varian Turbo-V 70D 969-9361 serie Manual De Instrucciones

Tabla de contenido

Publicidad

Connection A - HIGH VACUUM FLANGE

To connect the Turbo-V70D pump to the ISO inlet
flange, remove the outer ring and position the cen-
tering ring as shown in the figure.
Then fix the two flanges with the clamps or claws
as shown in the figure.

Connection configurations

For ConFlat flange connections we recommend
using Varian hardware.
To facilitate assembly and dismantling, apply Fel-
pro C-100 high temperature lubricant to the screw
threads protruding from the flange and between
the nuts and flange. Note that the 2.75î CF flange
has six blind holes.
Attach the units and tighten each one in turn. Re-
peat the sequential tightening until the flange
faces meet.
!
Exercise care when tightening nuts and bolts to
avoid creating dents in the envelope as this may
cause the pump rotor to lock.
35
TECHNICAL INFORMATION
CAUTION
87-900-861-01(E)

Publicidad

Tabla de contenido
loading

Tabla de contenido