Error Messages/Troubleshooting - Microlife BP3GR1-3P Manual De Instruccion

Monitor de presión arterial automatico
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Error messages/troubleshooting

If an error occurs during a measurement, the
measurement is discontinued and a corresponding
error code is displayed (example: Error no. 2).
Error No. Possible cause(s)/Solutions
ERR 1
ERR 2
ERR 3
ERR 5
HI
LO
*If this or any other problem occurs repeatedly, please
consult your doctor.
22
The tube may have loosened, or no pulse
was detected.* Ensure cuff connections
are tight with proper cuff placement. See
section 5.3.
Unnatural pressure impulses influenced
the measurement result. Reason: The arm
was moved during the measurement
(artefact). Repeat measurement, keeping
still and quiet.
Inflation of the cuff takes too long. The
cuff is not correctly seated or the hose
connection is not tight. Re-position cuff
and repeat the measurement.
The measured readings indicated an
unacceptable difference between systolic
and diastolic pressures. Take another
reading following directions carefully.
Contact your doctor if you continue to get
unusual readings.
The cuff pressure is too high. Relax for 5
minutes and repeat the measurement.*
The pulse is too low (less than 40). Repeat
the measurement.*

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido