Error Messages/Troubleshooting - Microlife BP3AQ-1 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6. Error messages/troubleshooting

If an error occurs during a measurement, the measurement is discontinued and
a corresponding error code is displayed (example: Error no.2).
Error No.
ERR 1
ERR 2
ERR 3
ERR 5
HI
LO
*If this or any other problem occurs repeatedly, please consult your
doctor.
Possible cause(s)/Solutions
The tube may have loosened, or no pulse was detected.*
Ensure cuff connections are tight with proper cuff placement.
See section 5.3.
Unnatural pressure impulses influenced the measurement
result. Reason: The arm was moved during the measurement
(artifact). Repeat measurement, keeping still and quiet.
Inflation of the cuff takes too long. The cuff is not correctly
seated or the hose connection is not tight. Re-position cuff
and repeat the measurement.
The measured readings indicated an unacceptable differ-
ence between systolic and diastolic pressures. Take
another reading following directions carefully. Contact
your doctor if you continue to get unusual readings.
The cuff pressure is too high. Relax for 5 minutes and
repeat the measurement.*
The pulse is too low (less than 40). Repeat the measurement.*
17

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido