TROUBLE SHOOTING
Problem
Display shows an
abnormal result
Display is blank
when power is on
No response when
POWER is pressed
or when new
batteries have been
loaded
Possible Cause
Cuff position was not correct
or was not tight enough
Body posture was not correct
during measurement
Speaking, excessive
movement, angry, excited
or nervous anxiety during
measurement
Batteries may be expired or
installed improperly
Incorrect operation; possible
strong electromagnetic
interference
Solution
Apply the cuff properly and attempt
a new measurement
Review the 'Taking a Measurement'
portion of this manual and attempt
a new measurement
Wait a period of time and attempt
a new measurement after calm has
returned; do not speak or move
during measurement
Check the polarity of the batteries
and reinstall if necessary; replace
batteries with new batteries
Remove batteries and wait a
minimum of five minutes; reinstall
batteries and attempt a new
measurement
ENGLISH • 23