Cleaning your Hybrigene
Before cleaning your unit ALWAYS disconnect from the power supply and allow
to cool.
You can clean the case of the Hybrigene with a cloth dampened in soapy water.
No part of the unit should be immersed in solvents. Do not use acetone or abrasive
cleaners.
DO NOT USE ETHANOL OR METHYLATED SPIRITS ON THE ACRYLIC
DOOR; these may make the acrylic go cloudy.
Before using any cleaning or decontamination method except those
recommended in this manual, the responsible body should check with Techne that the
proposed method will not damage the equipment.
If You Spill the Fluid
Because of the nature of operation of the Hybrigene, the complete inside must
be cleaned if a spillage does occur. This is to remove any airborne particles which
may have been blown around the system and to guarantee full operator safety.
Switch the Hybrigene off and disconnect it from the power supply. Follow this
sequence of operations when the Hybrigene has been allowed to cool below at least
50°C :
a)
Remove all glassware from the Hybrigene.
b)
Remove the Plastic drip tray.
c)
Clean all surfaces of the inner chamber using a suitable cleaning and
decontaminating solution.
Be careful of the drive shafts.
Cleaning and decontamination materials which Techne recommend include:
Lipsol from Bibby-Sterilin Ltd
Decon 90
Neutracon
or mild detergent.
22