Maintenance
The TwisTorr 305 FS does not require any maintenance. Any work must be
carried out by authorized personnel.
Before carrying out any work on the system, disconnect it from the power
WARNING!
supply, vent the pump with the vent valve, wait until the rotor is completely
stopped and wait until the surface temperature of the pump is below 50° C.
In the event the pump fails, you can use the Agilent repair service or
the "Agilent advanced exchange service", which gets you a remanufactured
system to replace the failed system. The exchange 305 FS is shipped when the
order is placed. You then return the failed 305 FS in the box that the exchange
305 FS comes in.
Before returning a pump for repair or advanced exchange service to the manufacturer,
NOTE
it is essential to complete and send to the local sales office the "Request for Return"
form attached to this instruction manual. Copy of the same must be inserted in the
packaging of the system before shipping.
If a system needs to be scrapped, ensure that local regulations are followed for
appropriate disposal.
Rotor life
The fatigue resistance of 305 FS turbomolecular pump rotor resulting from
theoretical evaluation is 405000 cycles of ramp up from stop to full speed and
back to stop.
As a precautionary measure, considering the typical usage of the pump and
taking into account safety limits, Agilent recommends to return the pump after
17 years of use for maintenance.
TwisTorr 305 FS User Manual / 87-901-057-01 (A.00)
Instructions for Use
Maintenance
207/258
7