Maintenance

Maintenance is not required unless the yellow LED is on. To check the error you
can read the code via serial connection. If a pump sits for a prolonged amount
of time, a visual check of the DB15 connection is recommended to make sure it
is free of oxidation.
Any other maintenance must be performed be authorized personnel.
Before carrying out any work on the system, disconnect it from the power
WARNING!
supply, vent the pump with the vent valve, wait until the rotor is completely
stopped and wait until the surface temperature of the pump is below 50° C.
In the event the pump fails, you can use the Agilent repair service or
the "Agilent advanced exchange service", which gets you a remanufactured
system to replace the failed system. The exchange 305-IC is shipped when the
order is placed. You then return the failed 305-IC in the box that the exchange
305-IC comes in.
Before returning a pump for repair or advanced exchange service to the manufacturer,
NOTE
it is essential to complete and send to the local sales office the "Request for Return"
form attached to this instruction manual. Copy of the same must be inserted in the
packaging of the system before shipping.
If a system needs to be scrapped, ensure that local regulations are followed for
appropriate disposal.

Rotor life

The fatigue resistance of 305-IC turbomolecular pump rotor resulting from
theoretical evaluation is 405000 cycles of ramp up from stop to full speed and
back to stop.
As a precautionary measure, considering the typical usage of the pump and
taking into account safety limits, Agilent recommends to return the pump
after 17 years of use for maintenance.