Interconnections; Earth (Ground) Connection - Agilent Technologies TwisTorr 305 FS Manual De Instrucciones

Tabla de contenido

Publicidad

Interconnections

The following figure shows the TwisTorr 305 FS interconnections.
Figure 25

Earth (Ground) Connection

We recommend that you fit a separate earth (ground) conductor to the
TwisTorr 305 FS. Use a separate insulated green/yellow conductor, and use
a M5 x 10 screw and shake proof washer (fitted to the earth hole on the pump) to
secure the earth conductor to the pump. The screw must be fixed with a 2 Nm torque.
The impedance between the pump-body and the earth connection point must
be < 0.1 Ω.
TwisTorr 305 FS User Manual / 87-901-057-01 (A.00)
Technical Information
EARTH GROUND
CONTROLLER
CONNECTOR

Interconnections

FORELINE
233/258
8

Publicidad

Tabla de contenido
loading

Tabla de contenido