MAINTENANCE

5.1

Error message

Low battery
Communication fail
Measured temperature out
of range (< -5° C or >55 °C)
Calibration out of range
Recall memory fail
The current data set has
already been stored once
Any more problems, you can contact Ohaus.
5.2

Meter maintenance

The STARTER 400D does not require any maintenance other than replacement of depleted
batteries. To clean, use a damp cloth.
Attention: Do not disassemble the instrument. There are no user serviceable parts inside.
The housing is made of acrylonitrile butadiene styrene (ABS). This material is susceptible to
damage by some organic solvents, such as toluene, xylene and methyl ethyl ketone (MEK). Any
spillage should be immediately wiped off.
Find Quality Products Online at:
If low battery, the battery icon will blink to show need
new battery. Then turn off after 10 seconds.
When do calibration or new cap parameter input, meter
will detect if the probe is connected properly. If not,
meter displays " Fail, please confirm the STDO21 probe
is connected"
Meter will displays "Temperature Out of Range!"
Please keep the sample temperature within the range for
measurement.
Meter displays "Calibration Fail".
|K-1| >0.4 or a |B| >0.1
You may need to do a proper calibration again, if
calibration data is still not good, you may need to replace
a new probe cap or probe.
Meter displays "No Data"
When recall memory while no data memorized yet.
Meter displays" Already saved!"
A measurement can only be stored once. Perform a new
measurement to store a new data set.