Gima OXY-100 Manual De Uso Y Mantenimiento página 28

Oxímetro
Ocultar thumbs Ver también para OXY-100:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

ENGLISH
28
4.4.1 Setting
On the menu screen, select "Setting" and then press "
" key for entering
into system setup screen. The setup screen is as shown in the following figures.
Figure 4-4 System Setup Screen
Operation Instructions:
- Patient: according to the type of SpO
probe, the matched patient type should
2
be selected, there are two options:"ADU" for adult, and "NEO" for neonate.
- SpO
Lo-Limit: SpO
low limit setting; range: 50%~99%, the step is 1%. The
2
2
factory default value for adult is 90% and 85% for neonate.
- PR Hi-Limit: High limit setting of pulse rate; range: 100~240bpm. From 100
to 150, the step is 1bpm, and from 150 to 240, the step is 5bpm. The factory
default value for adult is 120bpm and 160bpm for neonate.
- PR Lo-Limit: Low limit setting of pulse rate; range: 30~99bpm, and the step
is 1bpm. The factory default value for adult is 50bpm and 60bpm for neonate.
- Date: Date setting
When cursor stays on the Year of the date, press "
" (Mode/OK) key to
active Year option, the cursor flashes on the Year of the date;
Press "
/
" (Navigation key) to adjust year.
Press "
" (Power/back) key or "
" (Mode/OK) key to confirm and exit
from date setting.
The procedures of adjusting Month value and Day value are the same with
Year adjustment.
Date Format: yy-mm-dd
Note: The setting operations of other parameters (such as TIME, PATIENT,
RECORDING INTERVAL, POWER SAVING etc.) are the same with date
setting.