Purpose; Limitations Of Use; Application Restrictions; Operating Exclusions - BRAND Transferpette S-8 Instrucciones De Manejo

Ocultar thumbs Ver también para Transferpette S-8:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

leaks), stop pipetting immediately and refer to the section "Trou-
bleshooting"; see Troubleshooting, p. 58 . Contact the manufac-
turer, if necessary.

2.2 Purpose

This is an air displacement pipette for pipetting aqueous solutions of
medium density and low to medium viscosity.

2.3 Limitations of use

This instrument is intended for pipetting samples, within the following
limitations:
• Operating temperature of instrument and reagent should be be-
tween +15 °C and +40 °C (59 °F to 104 °F) (other temperatures upon
request)
• Vapor pressure up to 500 mbar
• Viscosity: 260 mPa s 260 mPa s
For viscous media, the speed must be adjusted if necessary.

2.4 Application restrictions

Viscous and wetting liquids may compromise volumetric accuracy. Volu-
metric accuracy may also be affected when pipetting liquids whose tem-
perature deviates from the ambient temperature by more than ± 1 °C/±
1.8 °F.

2.5 Operating exclusions

The user is responsible for checking the compatibility of the device with
the intended application. The device cannot be used:
• for liquids that corrode polypropylene and FKM
• for liquids that corrode polycarbonate
Transferpette® S-8/-12 997208
Operating manual
39

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Transferpette s-12

Tabla de contenido