Commands Specific To The Instrument "Multimeter Mode; Vertical - Chauvin Arnoux HANDSCOPE CA 922 Manual De Instrucciones

Tabla de contenido

Publicidad

37. COMMANDS SPECIFIC TO THE INSTRUMENT

37.1. VERTICAL

INPut{[1]|2|3|4}:DMM
:COUPling
INPUT{[1]|2|3|4}:DMM
:BANDwidth:AUTO
[SENSe]:RANGe
{[1]|2|3|4}:AUTO
[SENSe]:RANGe[1]:CAPA (Command/Query)
SENSe]:RANGe[1]:OHM (Command/Query)
[SENSe]:RANGe
{[1]|2|3|4} :VOLT
"MULTIMETER MODE"
(Command/Query)
The INP{[1]|2}:DMM:COUP <AC|DC|ACDC> command affects the coupling of the selected channel.
To the question INP{[1]|2}:DMM:COUP? the instrument returns the current coupling of the selected
channel.
(Command/Query)
The INP{[1]|2}:DMM:BAND:AUTO <1|0|ON|OFF> command limits the bandwidth of the channel to
5 kHz.
To the question INP{[1]|2}:DMM:BAND:AUTO? the instrument shows if the 5 kHz bandwidth limit is
active.
(Command/Query)
The RANG{[1]|2}:AUTO <1|0|ON|OFF> command authorizes or prohibits the autoranging of the selected
channel.
ON|1 activates the autoranging.
OFF|0 deactivates this function.
To the question RANG{[1]|2}:AUTO? the instrument returns the autoranging status for the selected
channel.
The RANG:CAPA <range|MAX|MIN|UP|DOWN> command selects the range of measurement to be
used in capacitance mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, the value is expressed in Farad.
To the question RANG:CAPA? the instrument returns the range value of the capacitance.
Response format:
value in format <NR3>
The RANG:OHM <range|MAX|MIN|UP|DOWN> command selects the measurement range to be used
in ohmmeter mode.
<range> is a value in format NRf, it may be followed or not by a multiple and by the unit.
By default, it is expressed in Ohm (Ω).
To the question RANG:OHM? the instrument returns the value of the measurement range of the
ohmmeter.
Response format:
value in format <NR3>
(Command/Query)
The RANG{[1]|2|3|4}:VOLT <range|MAX|MIN|UP|DOWN> command selects the measurement range
to be used in voltmeter mode for the selected channel.
<range> is a value in NRf format, it may be followed or not by a multiple and by the unit.
By default, it is expressed in volt.
To the question RANG{[1]|2|3|4}:VOLT? the instrument returns the value of the measurement range
of the voltmeter for the selected channel.
Response format:
value in format <NR3>
<range><NL>
<range><NL>
<range><NL>
93

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Handscope ca 942

Tabla de contenido