Ieee 488.2 Commands - Chauvin Arnoux HANDSCOPE CA 922 Manual De Instrucciones

Tabla de contenido

Publicidad

DDE
Device Dependant Error 3 (bit)
An error specific to the instrument has been detected.
QYE
Query Error (2 bit)
A query error has been detected.
RQC
Request Control (1bit)
Always at zero.
OPC
Operation Complete (0 bit)
All operations running are ended.
38.2.5. EVENT MASK REGISTER
Reading and writing  *ESE command.

38.3. IEEE 488.2 COMMANDS

*CLS
(Clear Status)
(Command)
The common command *CLS reset the status and event register.
*ESE
(Event Status Enable)
(Command/Query)
The *ESE <mask> common command positions the status of the event mask.
<mask> is a value in format <NR1>, from 0 to 255.
A 1 authorises the corresponding bit of the event register to generate an event, while a 0 masks it.
To the question *ESE?, the instrument returns the current content of the event mask register.
Response format:
value in format <NR1> from 0 to 255.
Event mask register:
*ESR?
(Event Status Register)
(Query)
To the question *ESR?, the instrument returns the content of the event register.
Once the register has been read, the content value is reset to zero.
Response format:
value in format <NR1> from 0 to 255.
Event register:
<value><NL>