Measurement Display - Chauvin Arnoux HANDSCOPE CA 922 Manual De Instrucciones

Tabla de contenido

Publicidad

MEASure:PERiod?
MEASure:FREQuency?
MEASure:PDUTycycle?
MEASure:PULse:COUNt? (Query)

36.5.3. MEASUREMENT DISPLAY

MEASure{[1]|2|3}:SELECT (Command/Query)
MEASure:AUTO
36.5.4. PHASE MEASUREMENT
MEASure:PHASe?
36.5.5. MANUAL MEASUREMENT
DISPlay[:WINDow]:CURSor
:STATe
(Query)
To the question MEAS:PERiod? <INT{1|2|3}> the instrument returns the period of the selected signal.
Response format:
value in format <NR3> expressed in second.
(Query)
To the question MEAS:FREQ? <INT{1|2|3}> the instrument returns the frequency of the selected signal.
Response format:
value in format <NR3> expressed in hertz.
(Query)
To the question MEAS:PDUT? <INT{1|2|3}> the instrument returns the duty cycle of the selected signal.
Response format:
value in format <NR2> expressed in percent.
To the question MEAS:PUL:COUN? <INT{1|2|3}> the instrument returns the pulse count on screen of
the selected signal.
Response format:
value in format <NR2>.
The command
MEAS{[1]|2|3} :SELECT <measure1>,<measure2> selects automatic measurements for display.
<measure1> our<measure2> can take values NO, MIN, MAX, PTPeak, LOW, HIGH, AMPLitude,
ROVERshoot, FOVERshoot, RTIME, FTIME, PWIDth, FWIDth, FREQuency, PERiod, PDUTycycle,
COUNt, RMS, AVG or PHASE.
To the question MEAS{[1]|2|3} :SELECT ? the instrument returns the current automatic measurements
selected for display.
Response format:
(Command/Query)
The command MEAS:AUTO <1|0|ON|OFF> activates the display of the selected automatic
measurements.
To the question MEAS:AUTO? the instrument returns the display activation state of the automatic
measurements.
(Query)
To the question MEAS:PHAS? <INT{1|2}> the instrument returns the phase of the first selected signal
to the second.
Response format:
value in format <NR2> expressed in degree.
(Command/Query)
The DISP:CURS:STAT <1|0|ON|OFF> command activates or inhibits the manual measurements.
1|ON: activates the manual measurements
0|OFF: inhibits the manual measurements
To the question DISP:CURS:STAT?, the instrument returns the activation status of the manual
measurements.
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measured value><NL>
<measure1>,<measure2><NL>
<measured value><NL>