BAR model
Err 8
9.2 WEIGHING UNIT LISTING
Nº
UNIT SYMBOL
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
10. TROUBLE SELF CHECKING
PROBLEM
Display is blank, No turn on
test
Display blank after turn on test,
Error message, Display is loc-
ked
OL or ---- appears on display
---- or NULL displayed
Display is unstable
external calibration weight error
g
g
ct
lb
oz
d
ozt
dwt
mom
tL.H
tl.c
tl.t
t
bt
n
POSIBLE CAUSE
Mains power is turned off.
Power supply not plugged in.
Power supply faulty.
Internal battery not charged.
Display turned off.
Pan not installed.
Unstable weight.
Load cell damaged.
Mechanics damaged.
Maximum capacity exceeded.
Load cell or mechanics damaged.
Power supply faulty.
Weight on scale below permissible li-
mit.
Pan has been removed.
Pan support not seated properly.
Power supply faulty.
Load cell or mechanics damaged.
Drafts or air currents.
Obstruction under pan.
Sample is moving (animal weighing).
Vibrations through table.
Temperature changed dramatically.
Power supply faulty.
More than 4% error as linear calibration
NAME OF THE UNIT
gram
gram (high resolution)
carat net
pound
ounce
drachm
ounce troy
scruples
momme
tael twn
tael chn
troy tael
tola
bangladeshi tola
newton
COMMON SOLUTIONS
Check power is going to the scale and
switch is on.
Verify the voltage going to the scale
matches the power supply labels on
the power module or scale.
Check the pans are installed correctly.
Try turning the scale on again.
Check the pans are installed correctly.
Try turning the scale on again.
Check the pans are installed correctly.
Try pressing
Try turning the scale on again.
Verify the scale is in a acceptable loca-
tion and on a good table.
Verify the power supply is correct for
the scale.
User's Guide
CONVERSION
= 0,2 g
= 453,59237 g
= 28, 349523000 g
= 1,7718450 g
= 31,1034768 g
= 1,55517400 g
= 3,749996 g
= 37,49996 g
= 37,799375 g
= 37,4290018 g
= 11,6638039 g
1bt= 11,6638039 g
1bt= 16Ana= 96Roti
1n= 102,04 g
key.
29