BAR model
8. PARAMETERS
The scale has 9 parameters that can be set by the user plus a method of entering the calibration section.
To set parameters press the
The display will show the first function, "F1 UNIT".
Pressing the
key will cycle through the other functions.
Pressing
key will allow you to set the function. Use the
Note: when display shows "TECH", press
key to enter technical parameters setting mode.
PARAMETERS TABLE
FUNCTION
Sets the which units want to use: g / ct / lb / oz / d / gn / ozt / dwt / mm / tl.T / tl.c / tl.t / t use
F1 Unt
to switch, use
Set the backlight to be on, automatic or off.
EL on: Backlight always on
EL Au: Backlight automatically turn on
F2 EL
EL off: Backlight allways off
Use
Set RS232
At first, select you will use RS232 or USB.
S 232: use RS232 interface
S USB: use USB interface
Then will select interface work mode
P Prt connect with mini printer
P Cont connect with remote display or PC
P Auto connect with mini printer, auto printer, auto print (auto accumulation)
wirELE: wireless, when you use RW-W with kit I or BW-W with kit I, you can plug wireless kit I to ba-
F3 Con
lance RS232 interface, balance will send data to RW-W or BW-W interface.
After you select communication mode, display will show b xxx, this is baud rate, you can select from
600, 1200, 2400, 4800 or 9600 bps.
If you select P Prt or P Auto (work with printer), after set baud rate, display will show tP or LP-50,
this is printer mode, you can select normal mini printer (tP) or label printer (LP-50).
If you select LP-50, display will show EnG or CH, this is print out language, you can select Chinese (CH)
or English (EnG).
Use
Pin
Enter the programming and calibration menus by entering the correct password (
tech
and press
Liner calibrate.
After you press
display will show LoAd 0, remove all weight from the pan and wait until display show next step.
Then display will show LoAd I, put the weight follow the display, wait for next step.
P1 Lin
Follow the display to load calibration weight on the pan.
Until display show LoAd F, load full capacity weight on the pan.
Note: calibration weight regulation: 0<load 1<load 2<load 3<load F(full capacity)
Load1, Load2, Load3 need be integer weight.
key during self checking.
key to enter, display shows "PIN", press
key to sure.
to switch, use
key to sure.
to switch, use
key to sure.
key to enter.
key, display will show Pin again, press
key to leave a parameter unchanged.
DESCRIPTION
,
,
User's Guide
,
,
key, and press
,
and
key and
key to enter,
)
27