Technical Data - ProMinent DULCOTEST CGE 2-mA-2 ppm Instrucciones De Servicio

Célula de medición para cloro orgánico
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Technical Data

14 Technical Data
Measured
variable
Free chlorine and organically bound chlorine
Area of
application
Drinking water, swimming pool water and water of a similar quality
Measuring range
CGE 2-mA-2 ppm
CGE 2-mA-10 ppm 0.1-10 mg/l
CGE 2-4P-10 ppm
pH range
5.5 - 9.5
Temperature
range
5 - 45 °C,
Temperature compensation
No sharp rises in temperature
Storage
temperature
Frost-free between 5 and 50 °C
Resolution
Corresponds with the value at the lower end of the measuring range
Max. pressure
DLG III A/B:
DGM:
Flow
DLG III and DGM in-line probe housing
Cross
sensitivity
Bromine, iodine, ozone, ClO
measurement errors
Conductivity
Test water
30 µS/cm - 10.000 µS/cm
Operating life
Membrane cap
Typically 1 year depending upon the water quality
Materials
Membrane cap:
Electrode adaptor:
Power supply
16 - 24 V DC; min 35 mA at 16 V DC
Output signal
4 - 20 mA (mA-Variant)
0 - 2 V
Enclosure rating
IP 65
32
0.02-2.0 mg/l
0.1-10 mg/l
1 bar (free flow)
3 bar (no pressure surges!)
, chloramines and other oxidising agents lead to
2
PPE
Electrode holder: PMMA
PVC
Electronics: Electronic components
(4P-Variant)
Optimum:
30 l/h
Minimum:
20 l/h
Maximum:
100 l/h
ProMinent
®

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Dulcotest cge 2-ma-10 ppmDulcotest cge 2-4p-10 ppm

Tabla de contenido