Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
Automated Cleaning without Cleaning Tools
_ Pre-cleaning
1
Remove contaminants on the surface with a cloth or soft bristle
brush under running tap water.
For products with moving parts, perform cleaning while moving
all parts (see "2 Moving Parts of Each Product").
≤ 38°C, ≥ 3.5 L/min, ≥ 30 sec.
Water quality: Drinking water
3
Place the product in the cleaning basket and set the cleaning basket in the washer-disinfector.
_ Cleaning and Disinfection
Perform cleaning and thermal disinfection using the washer-disinfector under the following conditions:
Temperature
Pre-cleaning
< 25°C
Cleaning
40 - 60°C
Rinsing
> 10°C
Thermal disinfection
90 - 93°C
Drying
Set the washer-disinfector according to the instruction of the manufacturer.
* Detergent used for validation:
Neutral enzymatic detergent: neodisher MediZym (DR. WEIGERT)
Alkaline detergent: neodisher SeptoClean (DR. WEIGERT)
_ Drying
1
Remove moisture on the surface of the product using a dry
cloth, air gun, etc.
2
Time
Water quality
3 min
Tap water
≥ 5 min
Distilled water
Deionized water
≥ 1 min
Demineralized water
5 min
2
17
Turn the Twist Collet to the closed position (if applicable).
Detergent
-
Neutral enzymatic detergent for medical
purposes (pH 7.0 - 8.0), or alkaline detergent
for medical purposes (pH ≤12)
-
-
Visually check that all contamination is removed.
In case there is residual contamination, repeat the procedures
of Pre-cleaning through Drying until all contamination is
removed.
Proceed to Lubrication (see "7 – 2 Lubrication")