Cleaning / Lubrication / Checking of the Attachment, Slim Attachment Hub and Slim Tube
1
Connect each part of PANA SPRAY Plus, EZ Spray Grip, and
each Spray Adaptor.
2
Mount the Slim Tube to the Slim Attachment Hub.
(for Slim Attachment 300 only)
3
Mount the Attachment to the Spray Adaptor.
4
Pull the product to make sure that it is securely connected.
● DO NOT use anything other than PANA SPRAY Plus for lubrication.
● Shake PANA SPRAY Plus well before use.
● Due to its structure, the lubrication oil of PANA SPRAY Plus comes out from the tip of the Attachment and the Twist Collet part. To
prevent dispersal of the lubrication oil in the surrounding area, it is recommended to cover the Attachment with a dry cloth or use the
Spray Mist Absorber.
● After lubrication, there could be high pressure inside the Spray Adaptor. When removing the Spray Adaptor, be careful because it may
pop out with force.
● Do not lubricate the Motor, Duraguard or Rotatable Duraguard. It may cause failure.
7 – 3 Checking
1
Confirm that there is no damage on the outer surface and
connection parts of the product or missing parts.
● To perform "Check before Operation", refer to OM-SH0913MA Primado2 OPERATION MANUAL Attachments.
5
Hold the Attachment firmly to prevent it from moving by the
pressure of PANA SPRAY Plus. Spray PANA SPRAY Plus for 3 - 5
sec.
6
Remove the Attachment from each Spray Adaptor by depressing
the Release Button Ⓐ on each Spray Adaptor.
7
Remove each Spray Adaptor from the EZ Spray Grip by
depressing the Release Button Ⓑ on the EZ Spray Grip.
Proceed to Checking (see "7 – 3 Checking").
CAUTION
2
Check with the same procedure described in "Check before
Operation".
Proceed to Packaging / Sterilization (see "10 Packaging /
Sterilization").
NOTICE
19