Preparation
3
3 – 1 Treatment at the Operating Site
1
Remove visible contaminants such as blood using a lint-free cloth moistened with distilled water, deionized water or mineral-free water
at the operating site.
2
Disconnect the Motor Cord, Foot Control Cord and Power Cord from the Control Unit.
3
Remove the Cutting Accessory, Motor, Hood, Beak, Duraguard, Rotatable Duraguard, Irrigation Nozzle and other connections from the
Attachment.
4
Remove the Slim Tube from the Slim Attachment Hub (for P300-1AHS/P300-1AHA only).
For removal procedure, refer to the Operation Manual of each product.
5