6.5.

Programming the control unit

6.5.1.

Changing the firing program

Indicator
on the
Main display
left
20 °C
PROG: 1
150°C/HR
632°C
Heating
zone display
Meaning
on the right
No program
in operation
Program number
selection
Indicator in the
segment display
Heating ramp
Cooling ramp
Soak/Target
temperature
62 / 317 - EN -
Description
If no firing is carried out, no element lights
up in the program sequence of the
graphical display.
The "Program running" indicator must not
light up
The main display shows the current firing
chamber temperature of the respective
heating zone.
g
When you press the
number lights up on the display.
The firing program can now be selected
d
f
with the
and
keys.
g
By pressing the
key again, the program
number which is to be changed can be
selected.
You can switch back to the previous value
in each step by pressing the
The selected firing program always shows the
first segment first.
The ramp rate in the main display appears
as:
⇒ "0,1°C/h-9,9°C/h" or
⇒ "10°C/h-999°C/h" or
⇒ "FULL" or
⇒ "END"
This can be altered with the
The graphical display indicates whether a
heating ramp or a cooling ramp is being
programmed.
Notes:
⇒ To change the heating ramp to a cooling
ramp and vice versa, see Section
6.5.2./6.5.3.
⇒ To program the ramp rate to "FULL" or
"END", see Section 6.5.4./6.5.5.
⇒ To program the additional switch outputs
1 & 2 see section 7.4.
The next press of the
soak or target temperature.
This can be altered with the
key, the program
h
key.
d
f
&
keys.
g
key displays the
d
f
&
keys.