Figure 3 Connection of feed tubes
Connection of the electrode
Electrical connection of the electrode
1.
Connect the electrode cable to the ELECTRODE port on the back
of the TitraLab.
The power is supplied via the TitraLab.
Connection of the electrode tubes
The sensor holder of the titration stand has two tube guides.
1.
Connect the tube from the OUT socket of the pump to the
electrode.
2.
To do this, push the free end of the tube into guide "1" of the sensor
holder until it reaches the middle of the electrode.
3.
Connect the tube on the OUT socket of the burette valve to the
electrode.
4.
To do this, push the free end of the tube into guide "2" of the sensor
holder until the end of the tube is flush with the tip of the electrode.
Figure 4 Connection of the electrode
Power supply, external
D A N G E R
Risk of electric shock
Make sure the AC supply voltage is suitable for the power supply and the mains
plug style is suitable for the type of outlet socket.
Only use the original power supply and plug connectors.
The power supply is provided with different plug connectors. The correct
plug connector must be attached to the power supply before first use.
English 83