Operation
Switch on the instrument
Become familiar with the instrument functions before operation. Learn
how to navigate through the menu and perform the relevant functions.
1.
Connect the input connector of the power supply to the POWER IN
port on the back of the TitraLab.
2.
Connect the power supply to a power socket.
3.
Press ON/OFF below the display to switch on the instrument.
4.
The instrument performs a self-test upon each activation.
Note: Wait approximately 20 seconds before switching on again so as
not to damage the electronics and mechanics of the instrument
Startup
Language and data output setting
The titrator software supports multiple languages. The first time the
instrument is switched on, a language selection list is displayed
automatically upon completion of the self-test.
1.
Select the required language.
2.
Press OK to confirm the selection.
3.
Select the type of data output (deactivated, printer, computer) and
confirm with OK.
Change the language setting
The instrument functions in the selected language until the option is
changed.
1.
Turn the instrument on.
2.
Select SYSTEM > LANGUAGE.
3.
Select the required language.
4.
Press OK to confirm the selection.
Chloride analysis in solid samples
1.
Place a magnetic stirring bar in a titration cell and dispense the
sample as follows:
Cheese or solid dairy products: Add 1 g of the sample dis-
solved in approximately 10 mL lukewarm distilled water
Tomato sauces: Add 1 g of the sample dissolved in approxi-
mately 10 mL of distilled water
Brine: Add 0,5 g of the sample dissolved in approximately
10 mL of distilled water
2.
Attach the titration cell to the magnetic stirrer.
3.
Before the analysis, stir the sample for at least 30 seconds in order
to extract the salt in full.
4.
Lower the electrode into the titration cell.
Note: Make sure that the electrode cannot be damaged by the
stirring bar or come into contact with the bottom of the titration cell.
Make sure the combined Silver rod electrode, including the
reference element porous pin is covered by sample.
5.
Select ANALYZE CHLORIDE (W). "SAMPLE: 1 WEIGHT" is
shown on the display.
6.
Enter the weight of the sample and confirm with OK.
The following steps are carried out automatically by TitraLab:
Acid is dispensed into the titration cell for 2 seconds
The solution is stirred for 3 seconds
Titration is then started
7.
The result is shown on the display and, depending on the setting,
sent to a printer or computer.
8.
Press OK to confirm the result.
9.
To begin another analysis, select OK. "NEXT SAMPLE" is shown
on the display.
English 85