B.A. International BA45 Instrucciones De Uso página 49

Ocultar thumbs Ver también para BA45:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

6453505.book Seite 47 Donnerstag, 21. April 2016 4:59 16
Operating Instructions BA ULTIMATE Kit
7 Care and maintenance
The cleaning and disinfection equipment used must be approved by its
Cleaning and disinfection equipment
manufacturer for the cleaning and disinfection of dental instruments and
comply with EN ISO 15883-1 (e.g. 95°C (203°F) and 10 min. holding
time).
For further details, refer to the operating instructions supplied with the
unit.
1. Check whether BA ULTIMATE is clean after processing.
Check _ product name
2. If it is still dirty, repeat the process.
 For further processing, BA ULTIMATE should be dry and free of
residue.
3. Blow BA ULTIMATE out with max. 3 bar.
4. Manual care of mechanical parts [ → 37].
5. Manual care of the push button chuck [ → 43].
47
64 53 505 D3551.201.04.02.02

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Ba175Ba605Ba606Ba695fbBa695fm

Tabla de contenido