7 Care and maintenance
134°C
274°F
48
7.5
Sterilizing
1. Clean and disinfect BA ULTIMATE.
Sterilizing
Steps _ product name
2. Care BA ULTIMATE with spray [ → 37].
3. Sterilize BA ULTIMATE in the steam sterilizer with saturated water
vapor.
Saturated water vapor:
Temperature: 134°C (274°F)
Overpressure: 2.04 bar (29.59 psi)
Holding time:
3 min
Steam sterilizers that meet the requirements of EN 13060, class B or S,
without DAC PROFESSIONAL_Instrument
and are also suitable for the sterilization of straight/contra-angle
handpieces are approved.
NOTICE! Do not exceed 140°C (284°F), even during the drying phase.
Operating Instructions BA ULTIMATE Kit