Operating Instructions BA ULTIMATE Kit
64 53 505 D3551.201.04.02.02
BA ULTIMATE can be sterilized in packaging that is suitable for
Product name
In-packaging sterilization
sterilization and storage, e.g. paper/plastic composite packaging.
After sterilizing
After sterilizing - heading
1. Remove BA ULTIMATE from the steam sterilizer immediately.
Follow-up _ product name
CAUTION! BA ULTIMATE is hot.
NOTICE! Do not try to accelerate the cool-down process by placing
it in cold water. This will damage your BA ULTIMATE!
2. Store BA ULTIMATE so that it is protected against contamination.
3. Sterilize again once the storage period has elapsed.
7.6
Cleaning the cooling spray nozzle
If your tap water is very hard, calcareous deposits can constrict or
Cleaning the cooling spray nozzle
completely block the cooling spray nozzle.
1. Carefully clean the nozzle opening on a regular basis by running a
Cleaning the nozzles _ Turbine
cleaning wire through it.
7 Care and maintenance