4 Operation
b.
c.
d.

4.6 Adapter for 25 ml and 50 ml

Adapter
Operating manual
46
a.
Hold the HandyStep® S over a vessel.
b.
Empty the PD-Tip by pushing the
locking/filling lever to its lower stop.
c.
Swing out the locking/filling lever
(this will cause the piston to decou-
ple).
d.
Hold the HandyStep® S over a waste
container. Press the ejector button
down (releases the cylinder).
ð PD-Tip is ejected.
To use the 25 ml and 50 ml PD-Tips, the
reusable adapter supplied with the tips is
required. This is coupled to the PD-Tips
with a bayonet lock and can be detached
after use. Adapters are autoclavable at
121 °C (2 bar), in accordance with DIN EN
285.