Error Message
Malfunction (Info)
[Mixing motor error:
[Power consumption
Bearing current]
too high]
[Mixing motor error:
[Impeller is
Impeller unstable]
unstable]
[Mixing motor error:
[Open drive coil]
Drive coil]
[Mixing motor error:
[Controller power
Power output]
output short circuit
or overtemperature]
[Mixing motor error:
[The rpm of the
Speed lost]
impeller is 0]
[Mixing motor error:
[No impeller in the
No impeller
drive – either not
detected.]
present or too far
away.]
Cause
Solution
The storage current
Check the position of the
is too high, causing
mixing bag and impeller.
the mixing process
to be interrupted or
locked.
The impeller in the
Check the position of the
mixing bag is
mixing bag and impeller.
unstable, causing
the mixing process
to be interrupted or
locked.
The Motor has a
Switch off the device.
fault, causing the
Check whether the power
mixing process to be
supply cable is properly
interrupted and
connected.
locked.
Switch on the device.
The listed output is
Switch the device off and on
exceeded, causing
again.
the mixing process
to be interrupted
and locked.
The mixing speed
Stop the mixing process.
decreases to 0
during the mixing
Decouple the device from the
process and does
impeller and Palletank
not increase again
mixing.
within 10 seconds.
Couple the device with the
impeller and Palletank
mixing.
The impeller is not
Check the position of the
recognized.
mixing bag and impeller.