Malfunctions
10.3 Faults when Operating the Device
Fault
The device will not turn on.
The program menu or the
administration menu are not
displayed.
10.4 Leaked Liquids
Fault
The device has come into
contact with liquid because
large amounts of liquids
have leaked out of the
Palletank
for Mixing.
®
10.5 Checking the Position of the Mixing Bag and Impeller
Procedure
Decouple the device from the impeller and Palletank
t
Check that the bag and the impeller are correctly installed in the Palletank
t
Palletank
for Mixing).
®
If required: Adjust the position of the mixing bag.
t
Insert the device under the Palletank
t
Page 40 and Chapter "8.6", Page 42).
If the warning message reoccurs: Contact Sartorius Service.
t
10.6 Checking the Coupling with the Mixing Bag and Impeller
Procedure
Stop the mixing process.
t
Decouple the device from the impeller and Palletank
t
Wipe the Palletank
recognition sensor on the device with a damp cloth.
®
t
Insert the device under the Palletank
t
"10.5", Page 52 and Chapter "8.6", Page 42).
52
Operating Instructions Flexsafe
Cause
The emergency shut-off
switch is engaged.
The necessary authorizations
have not been assigned to
the personal user profile.
Cause
The mixing bag is damaged
and has fully or partially
emptied. The liquid may
have penetrated into the
device via the fan.
for Mixing (see Chapter "8.12.1", Page 45).
®
for Mixing and couple with the mixing bag's impeller (see Chapter "8.5",
®
for Mixing (see Chapter Chapter "8.12.1", Page 45).
®
for Mixing and couple with the mixing bag's impeller (see see Chapter
®
Pro Mixer Drive Unit
®
Correction
Disengage the emergency shut-off
switch. To do so, pull out the
emergency shut-off switch until the
green ring on the emergency shut-
off switch is visible.
Check the authorizations of the user
profile.
Correction
Contact Sartorius Service.
for Mixing (see instructions for the
®
Chapter,
page
13.1,
55
Chapter,
page