Accuracy
Refer to the "Specifications" section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount) at 23 °C ± 5 °C
(73.4 °F ± 9 °F), 0% to 75% relative humidity.
True RMS Readings: DM-510A AC accuracies are specified from 5% to 100% of the range unless
otherwise specified. Frequency must be within the specified bandwidth for non-sinusoidal waveforms.
Crest factors are as follows:
Crest factor < 3:1 at full scale
Crest factor < 6:1 at half scale

All Models

AC Voltage
Range
(50 Hz to 400 Hz)
60.00 mV
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
Input Impedance: 10 MΩ // 50 pF
Accuracies for DM-510A are specified from
5% to 100% of range
Resistance
Range
600.0 Ω
6.000 kΩ
60.00 kΩ
600.0 kΩ
6.000 MΩ
60.00 MΩ
Open Circuit Voltage: 0.45 VDC typical
Accuracy
± (1.0% + 0.05 mV)
± (1.0% + 0.5 mV)
± (1.0% + 0.005 V)
± (1.0% + 0.05 V)
± (1.0% + 0.5 V)
± (1.0% + 5 V)
Accuracy
± (0.5% + 0.4 Ω)
± (0.5% + 0.004 kΩ)
± (0.5% + 0.04 kΩ)
± (0.5% + 0.4k Ω)
± (0.7% + 0.004 MΩ)
± (1.2% + 0.04 MΩ)
DM-200A • DM-210A • DM-510A
DC Voltage
Range
60.00 mV
600.0 mV
6.000 V
60.00 V
600.0 V
1000 V
Input Impedance: 10 MΩ // 50 pF
Diode Test
Measuring Range: 1.000 V
Test Current (typical): 0.56 mA
Open Circuit Voltage: < 1.8 VDC
Accuracy: ± (1.0% + 0.003 V)
Continuity
Tone Threshold: Between 10 Ω and 80 Ω
Response Time: < 32 ms
Accuracy
± (0.4% + 0.05 mV)
± (0.2% + 0.3 mV)
± (0.2% + 0.003 V)
± (0.2% + 0.03 V)
± (0.2% + 0.3 V)
± (0.2% + 3 V)
17