Electrical connection of the magnetic stirrer
1.
Connect the cable of the magnetic stirrer to the STIRRER port on
the back of the TitraLab.
The power is supplied via the TitraLab.
Installation of tubes
WA R N I N G
Risk of injury to eyes, skin and respiratory
During the first days of system use, check the screw joint between the syringe
and valve to avoid leaks and the potential escape of reagents.
When working with chemicals and/or solvents, always observe the relevant acci-
dent prevention regulations and wear appropriate personal protective clothing for
eye, face, hand, body and/or respiratory protection.
Make sure that the tubes are never kinked.
Regularly check the system as well as all tubes and connections for
leaks, condition and cleanliness.
To prevent twisting of the tubes, you should always connect flange tube
connections before those with grommets.
Removal of protective stickers (Uninstalled tubes
configuration)
1.
Remove the protective stickers on the burette valve and the four
pump connections.
2.
Make sure that the pre-installed white O-rings are not removed.
3.
Check that the O-rings are in a horizontal position inside the thread.
English 86
Figure 4 Position of protective stickers
Connection of feed tubes
The system is delivered with three feed tubes with bottle closures (DIN
45 thread). These feed tubes are supplied pre-installed on the TitraLab
and only need to be connected to the relevant reagent bottles.
1.
Connect the tube from the burette valve to the titrant bottle. Screw
onto the bottle closure to do this.
2.
Connect the tube from the IN socket of the ACID pump to the acid
bottle. Screw on the bottle closure to do this.
3.
Connect the tube from the IN socket of the ALKALI pump to the
alkali bottle. Screw on the bottle closure to do this.
4.
Place a desiccant cartridge in each bottle closure.