HBS 300i x-touch
8.3 Foot switch settings
After connecting the foot switch on the back, the "Auto-detection"-Symbol appears on the screen. The
symbol indicates which foot switch was connected.
It is important to ensure that the instruments are connected to the connectors, which are selected by
footswitch. For example: In Monopolar 1 connected instruments can only be operated with the
settings MONO1 of the foot switch.
Example:
The "auto-detection"-Symbol is displayed on
Monopolar2, which means that the
foot switch is connected to "Mono2" on the
back of the unit.
8.4 Handles with buttons
Activating the handle with buttons takes place independently of the foot switch settings. These cannot
be activated using the foot switch.
8.5 Neutral electrode settings and attachment
In the monopolar operating mode, a neutral electrode is required.
After attaching the neutral electrode to the patient, the HF unit recognize the neutral
electrode automatically (cf. illustration on the left and chapter 5.2 "
operating modes and settings
divided and undivided contact surfaces.
If the displayed neutral electrode type does not agree with the used
neutral electrode type, this is indicative of poor adhesion of the
neutral electrode on the patient, a defective neutral electrode or a
Please note the following information when attaching the
"). The HBS 300i x-touch recognizes neutral electrodes with
defective cable!
neutral electrode
EN
Overview of symbols for
45