8.11 Operating errors
In the case of an operating error, the following display appears:
The following error messages can occur:
Error - display
Cause of error
Neutral electrode not connected,
NE Error - No 1
not known, or NE is defective
Cut button on handle in terminal
Mono1 - Cut hand
Mono 1
Mono1 - Coag
Coag button on handle in terminal
hand
Mono 1
Cut button on handle in terminal
Mono2 - Cut hand
Mono 2
Mono2 - Coag
Coag button on handle in terminal
hand
Mono 2
If one or more numerical codes are displayed, or if the device is indicating any other recognizable
defects, please refer to service@hebumedical.de.
◼
HF current does not flow when the foot switch is actuated. The system only reacts with a
tone signal and the display changes colour.
Please check:
➔
That all terminals have been correctly connected
➔
The settings of the foot switch
➔
Whether the neutral electrode has been recognized
◼
The neutral electrode is not or not correctly detected
Check all terminals and touch the field with the symbol of the neutral electrode for detection of the
NE. If the neutral electrode is still not recognized, exchange it where appropriate.
◼
It is not possible to switch the machine on or off.
Check whether the power cable has been correctly connected and that the socket is intact. If the
error persists, inform the specialist personnel.
◼
The system works perfectly, but the settings cannot be saved.
Inform the specialist personnel.
Remedying errors
Check the terminal, touch the NE symbol
on the touchscreen for detection, where
applicable exchange NE and/or
connecting cable.
Check the terminals, check the settings,
exchange accessories if necessary. If
error persists, inform specialist personnel.