Descargar Imprimir esta página

Nouvag Vacuson 60 LP Instrucciones De Uso página 50

Ocultar thumbs Ver también para Vacuson 60 LP:

Publicidad

Idiomas disponibles

Idiomas disponibles

Packaging
Reassemble the Conform Cannula Handpiece (view 6.5 Preparing Conform Cannula Handpiece).
Individually: Pack Conform Cannula Handpiece in an individual packaging for sterile items.
Sets:
on
Sterilization
Autoclave in the vacuum autoclave (Class B or S as per EN 13060 with fractionated pre-vacuum)
at 135°C for at least 5 minutes. When sterilizing several instruments in one sterilization cycle, do
not exceed the sterilizer's maximum load. A drying cycle must be added in the case of auto-
claves without a post-vacuum function. Dry the Conform Cannula Handpiece in the bag with the
paper side facing upwards at room temperature for at least one hour
* The temperature holding times are based on country specific guidelines and standards.
Maximal exposure time: 25 min.
Storage
If the sterilized Conform Cannula Handpiece is not used immediately following sterilization, the
packaging needs to be marked with the sterilization date. It is advisable to add a sterility indica-
tor.
The above-indicated instructions were validated by Nouvag AG as being suitable for preparing a medical device to be reused.
The preparer is responsible for ensuring that the actually performed preparation involving the equipment, materials and personnel used in the
preparation facility achieves the desired results. This normally requires validation and routine monitoring of the procedure.
Likewise, each deviation from the provided instructions should carefully analyzed for its effect and potential detrimental consequences.
8.9
Electronicmotor 21, reprocessing instruction
For the reprocessing instructions for the electronic motor, please refer to the operation instructions sup-
plied with the electronic motor.
20
The bag must be large enough to ensure that the seal is not under tension.
Nouvag AG recommends to add a sterility indicator.
Sort Conform Cannula Handpieces onto trays intended for the purpose or place
all-purpose sterilization trays.
Operating instructions Vacuson 60 LP, REF 31793, V34/17
Vacuson 60 LP

Publicidad

loading