– Changing the range manually will also restart the recording session.
NOTE
– You can also use the MaxMin function while measuring frequency (see
– If an overload is recorded, the averanging function will be stopped. OL is
– The APO (auto power-off) function is disabled when MaxMin is enabled.
– The maximum recording time is 99999 seconds (1 day, 3 hours, 46 minutes,
4 Press
This mode is useful for capturing intermittent readings, recording minimum and
maximum readings unattended, or recording readings while equipment operation
keeps you from observing the multimeter display.
The true average value displayed is the arithmetic mean of all readings taken
since the start of recording. The average reading is useful for smoothing out
unstable inputs, calculating power consumption, or estimating the percentage of
time a circuit is active.
Keysight U1273A/U1273AX User's Guide
"Measuring Frequency"
reflected accurately, press
shown in place of the average value.
39 seconds). OL is shown if the recording exceeds the maximum time.
or
for more than 1 second to disable the MaxMin function.
Dual
Exit
on page 100). If the measured frequency shown is not
again to restart the recording session.
Dual
Exit
Multimeter Features
3
111