Multimeter Setup Options
4
Enabling and disabling the Filter
There are two Filter (
) options within the design of the multimeter:
– In AC/AC+DC measurement mode, the Filter works as a low pass filter and
attenuates signals with frequencies of more than 1 kHz
– In DC measurement mode, the Filter will block AC signals
Only one of the two filters can be in the signal path at any point in time. The
possible scenarios are:
– Only the AC Low Pass Filter is enabled
– Only the DC Filter is enabled
– No Filter in the signal path
The
icon appears when either of the filter circuits are enabled. When either
LPF is enabled, the measurement speed (response time) will be impacted.
The DC Filter cannot be used when the dual display mode is enabled where AC
NOTE
and DC voltages are measured.
Keysight U1271A/U1272A User's Guide
131