Descargar Imprimir esta página

Ivoclar Vivadent Bluephase PowerCure Instrucciones De Uso página 123

Ocultar thumbs Ver también para Bluephase PowerCure:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 82
Guidance and manufacturer's declaration – electromagnetic immunity
Bluephase PowerCure is intended for use in the electromagnetic environment specified below. The customer or the user of
Bluephase PowerCure should assure that it is used in such an environment.
Immunity test
IEC 60601 test level Compliance level
Conducted RF
3 V
rms
IEC 61000-4-6
150 kHz to 80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.5 GHz
Table: According to IEC60601-1-2
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio
a
broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which Bluephase PowerCure is used exceeds the applicable
RF compliance level above, Bluephase PowerCure should be observed to verify normal operation. If abnormal performance is ob-served, additional measures may
be necessary, such as reorienting or relocating Bluephase PowerCure.
Over the frequency range 150 kHz to 80 MHz, field strength should be less than 10 V/m.NOTE: U
b
Recommended separation distances between portable and mobile RF communications equipment
and "Bluephase PowerCure"
Bluephase PowerCure is intended for use in the electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of Bluephase PowerCure can help prevent electromagnetic interference by maintaining a miminum distance
between portable and mobile RF communications equipment (transmitters) and Bluephase PowerCure as recommended below,
according to the maximum out-put power of the communication equipment.
Rated maximum output
power of transmitter
(W)
0.01
0. 1
1
10
100
Table: According to IEC60601-1-2
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be determined using the equation
applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 4: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
10 V
10 V/m
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 0.35 √P
0.035
0. 1 1
0.35
1. 1
3.5
Electromagnetic environment – guidance
Portable and mobile RF communications equipment
should not be used closer to any part of Bluephase
PowerCure, including cables, than the recommended
separation distance caculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d = 0.35 √P
d = 0.35 √P 80 MHz to 800 MHz
d = 0.70 √P 800 MHz to 2.5 GHz
Where P is the maximum output power rating of the
trans mitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined
by an electro magnetic site survey,
the compliance level in each frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
is the a.c. mains voltage prior to application of the test level.
T
80 MHz to 800 MHz
d = 0.35 √P
0. 1 2
0.38
1.2
3.8
12
should be less than
a
b
800 MHz to 2.5 GHz
d = 0.7 √P
0.23
0.73
2.3
7.3
23
123

Publicidad

loading