Descargar Imprimir esta página

Ivoclar Vivadent Bluephase PowerCure Instrucciones De Uso página 6

Ocultar thumbs Ver también para Bluephase PowerCure:

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 82
Product Overview
1.4
Operating the curing light
Program selection button
to select the desired
curing program
Activating/deactivating polyvision
The polyvision function is activated by long pressing (>2 s) the program or time selection button
(see 3.2 Operation). To confirm that polyvision is activated, the handpiece will beep and vibrate
for a short time and the polyvision indicator will light up. Renewed long pressing of the program
or time selection button results in the polyvision function being deactivated; the handpiece will
not vibrate.
With the handpiece switched on, the current charging status is
shown on the handpiece as follows:
• No indicators lit up on the handpiece: Battery sufficiently
charged
Curing capacity of minimum 20 minutes in the High Power
program, approx. 15 minutes in the Turbo program and
approx. 8 minutes in the 3sCure program.
• Battery symbol on the handpiece is flashing orange:
Battery weak
Time/intensity can still be set and a polymerization time of
approx. 6 minutes in the High Power program, approx. 4 minutes
in the Turbo program, or 30 cycles in the 3sCure program are left.
Place the device into the charging base as soon as possible.
• Battery symbol on the handpiece is flashing orange
and a red "x" is shown: Battery completely discharged
The light can no longer be activated and the curing time can
no longer be set. However, the handpiece can be used in the
Click & Cure corded mode.
6
Start/Stop button
to trigger/abort the
curing process
Time selection button
to select the desired
curing time

Publicidad

loading