16. Technical results
EMC requirements
EN
IEC 60601-1-2 Edition 4.0 (2014)
Environment of intended uses:
Home Healthcare Facility Environment
Summary of Test Results
Emission (IEC 60601-1-2 section 7)
Radiated emission
Freq. range: 30 - 1000
MHz
Immunity (IEC 60601-1-2 section 8)
Immunity from
Electrostatic discharge
(ESD)
Immunity from radiated
electromagnetic fields
Immunity from Proximity
field from wireless
communications
equipment
Immunity from power
frequency magnetic field
Simplified Declaration of Conformity: In accordance with article
10(9) of the EU declaration of conformity, the radio equipment
type Bluetooth LE is in compliance with Directive 2014/53/EU.
For the complete document go to: https://silkn.co.uk/
media/downloads/EU-Declaration-of- Conformit y.pdf ?_
ga=2.228140826.664266296.1655719376-935110399.1646405517
Test
Standard
CISPR 11
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-3
IEC 61000-4-8
Class/ Severity level
Group 1 Class B
8 kV contact discharges &
15 kV air discharges
3.0; 10.0 V/m; 80 MHz
GHz,
80% AM, 1 kHz
List of frequencies,
from 9 V/m up to 28 V/m,
PM (18 Hz or 217 Hz), FM 1 kHz
30 A/m @ 50 Hz
28
Test
result
Complies
Complies
Complies
2.7
Complies
Complies

Publicidad