28

8. REMOTE CONTROL OF INSTRUMENT

The instrument can be remote controlled via ASCII codes and the RS232 port.
-
The RS232 port has to be configured according to OPTO RS (see Chapter 9).
-
Up to 64 control commands can be sent at the same time.
-
It is not possible to access keyboard functions whenever a sequence of control commands is
being executed.
-
Upon occurrence of the first error, all subsequent control commands will be cancelled.
-
When an error occurs, further control commands are not accepted with the exception of:
"ERR?","!ERR" and "RST".

8.1 Control command attributes

The execution of certain control commands is only possible with the instrument in a specific
configuration (control command attributes). An error message is initiated whenever a control
command is sent where compliance is not given for a required attribute.
List of control command attributes:
1.
ATTR_REF => The reference has to be adopted.
2.
ATTR_CONST => The probe constant must have been measured.

8.2 List of control commands

Attribute
1, 2
1, 2
1, 2
1
1
1
1
1
1
1
1
Control
Description
command
PRI
Used to print the currently indicated value.
?
Used to print the currently indicated value.
CLE
Used to reset the minimum and maximum values.
ID?
Used to print the instrument values.
IN
Used to set the current unit to inches.
MM
Used to set the current unit to millimetres.
KEY0
Used to block all keys of the keyboard (alias of: LCK0123456789.E).
KEY1
Used to unlock all keys of the keyboard (alias of: ULK0123456789.E).
AXI
Used to set the current mode to "Centre distance".
DIA
Used to set the current mode to "Diameter".
DEL
Used to set the current mode to "Delta".
NOR
Used to set the current mode to "Normal" (height).
MAX
Used to set the current mode to "Max".
MIN
Used to set the current mode to "Min".
MOD?
Used to print the current mode ("NOR", "DIA", "AXI", "MIN", "MAX" or "DEL")