Operation
5.2
Check of parameters
To ensure that the values were correctly transferred to the gas detection
instrument:
1.
Press the data from X-am 1/2/5x00 in Dräger CC Vision.
2.
Check parameters.
6
Operation
6.1
Preparation for operation
Before using the instrument for the first time, insert a charged NiMH T4
power pack or batteries approved by Dräger, see Chapter "Replacing the
batteries / rechargeable batteries" on Page 41.
The instrument is ready for operation
!
WARNING
To reduce the risk of ignition of a flammable or explosive atmosphere, strictly
adhere to the following warning statements:
Only use power pack types ABT 01xx, HBT 00xx or HBT 01xx. See the marking
on the rechargeable battery for permitted rechargeable batteries and the
corresponding temperature class.
Substitution of components may impair intrinsic safety.
6.2
Switching on the instrument
1.
Hold down the [OK] key for approx. 3 seconds until the countdown
» 3 . 2 . 1 « shown on the display has elapsed.
All the display segments, including the visual, audible and vibration
alarms, are activated for a short time.
The software version is displayed.
The instrument performs a self-test.
The sensor that is up next for adjustment is displayed with the
remaining days until the next adjustment, e.g. » Ex %LEL CAL 20 «.
The time until the bump test interval elapses is displayed in days, e.g.
» bt 123 «.
All alarm thresholds A1 and A2 as well as
or all toxic gases (e.g. H
1)
Only when activated in the instrument configuration. Delivery condition: not activated.
34
S or CO) are displayed consecutively.
2
1)
(TWA)
and
1)
(STEL)