Descargar Imprimir esta página

Dräger X-am 5000 Serie Instrucciones De Uso página 40

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 76
en
Operation
6.7
Info Mode
6.7.1
Activating the Info mode
In measuring mode, press the [OK] key for approx. 3 seconds.
If any warning or fault messages exist, the corresponding note or error
codes are displayed (see Technical Handbook).
Press the [OK] key successively for the next display. The peak values and
the exposure values TWA and STEV will be displayed.
If no key is pressed for 10 seconds, the instrument returns automatically to
measuring mode.
6.7.2
Info Off mode
Press the [ + ] key when the instrument is turned off.
The name of the gas, measuring unit, and measuring range limit value are
displayed for all channels.
Press the [ + ] key again to exit the Info Off Mode (or via timeout).
6.7.3
Calling the Quick Menu
In measuring mode, press the [ + ] key three times.
If functions in the quick menu are activated using the Dräger CC Vision PC
software, you can select these functions using the [ + ] key. If no functions
have been activated in the quick menu, the instrument remains in measuring
mode.
Possible functions:
[OK] key to retrieve the retrieve the selected function.
Press the [ + ] key to cancel the active function and to switch to measuring
mode.
If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
40
1. Bump test (configuration for bump test,
see technical manual)
2. Fresh gas adjustment
3. Delete peak values
To display pump information,
see the Technical Manual
To activate or deactivate the pump,
see the Technical Manual

Publicidad

loading

Este manual también es adecuado para:

X-am 5600