SCPI Command
[SOURce[1|2]:]SWEep:HTIMe {<hold_time>|MINimum|MAXimum|DEFault}
[SOURce[1|2]:]SWEep:RTIMe {<return_time>|MINimum|MAXimum|DEFault}
Marker Frequency
If desired, you can set the frequency at which the signal on the front panel Sync Out connector goes to a logic low
during the sweep. The Sync signal always goes from low to high at the beginning of the sweep.
– Marker frequency: 1 μHz to maximum frequency for the waveform. The default is 500 Hz.
– When the sweep mode is enabled, the marker frequency must be between the specified start frequency and stop
frequency. If you attempt to set the marker frequency to a frequency not in this range, the instrument will set the
marker frequency equal to the start frequency or stop frequency (whichever is closer).
– You cannot configure the marker frequency with the front panel menus unless the Sync source is the sweeping
channel.
Front Panel Operations
1. Press [Sweep] > Sweep ON |
2. Press [Trigger] > Sync
3. Select Mode
Marker
124
OFF
> Sweep
ON
ON
| OFF > Sync Setup .
.
| OFF .
Keysight EDU33210 Series User's Guide