SCPI Command
[SOURce[1|2]:]BURSt:NCYCles {<num_cycles>|INFinity|MINimum|MAXimum}
Burst Period
Burst period, which is used in internal triggered burst mode only, is the time from the start of one burst to the start
of next burst (1 μs to 8000 s, default 10 ms). Burst period differs from "waveform frequency," which specifies the
frequency of the bursted signal.
Burst period is used only when Immediate triggering is enabled. The burst period is ignored when manual or
external triggering is enabled (or when the gated burst mode is selected).
You cannot specify a burst period that is too short for the instrument to output with the specified burst count and
frequency. If the burst period is too short, the instrument will increase it as needed to continuously re-trigger the
burst.
Front Panel Operations
Press [Burst] > Burst Period . Use the numeric keypad or the knob and arrows to set a desired value. If you use the
keypad, select a prefix unit to finish.
132
Keysight EDU33210 Series User's Guide