SCPI Command
[SOURce[1|2]:]BURSt:INTernal:PERiod {<seconds>|MINimum|MAXimum}
Start Phase
Start phase of the burst, from -360 to +360 degrees (default 0).
– Specify the start phase units with UNIT:ANGLe.
– Always displayed in degrees on front panel (never radians). If set in radians from remote interface, instrument
converts value to degrees on the front panel.
– For sine, square, and ramp, 0 degrees is the point at which the waveform crosses 0 V(or DC offset) in a positive
going direction. For arbitrary waveforms, 0 degrees is the first waveform point. Start phase has no effect on noise.
– Start phase also used in gated burst mode. When the gate signal goes false, the current waveform cycle finishes,
and output remains at the voltage level of the starting burst phase.
Front Panel Operations
Press [Burst] > Start Phase . Use the numeric keypad or the knob and arrows to set a desired value. If you use the
keypad, press Degrees to finish.
Keysight EDU33210 Series User's Guide