1 Introduction
An electrosurgical unit is a generator used to transform electronic energy into high-frequency
alternating current (HF current).
If this current flows through human tissue, it creates heat which is used for cutting and for coagulation
of the tissue.
The HBS x-touch 300 is a versatile and compact electrosurgical unit which complies to a high safety
standard and unites economy with operative flexibility.
It offers the following application modes:
For monopolar surgical use
Pure cut
Blend cut
Polypectomy mode
Cutting in water environment (TUR)
Soft coagulation
Forced coagulation
Spray coagulation
For bipolar use:
Blend cut
Bipolar Coagulation
Bipolar Cutting in water environment (TUR)
Benefits of the HBS x-touch 300:
Simple, intuitive and safe operation
USB-connection for service
Current terminal for optional wireless foot switch, max. 0.5 A
Display-supported neutral electrode safety system helps interactively when applying the neutral
electrode
Nine storage locations for individual program settings
Two monopolar multifunction terminals for 3-pin standard or Martin plug
AutoStop function for automatic interruption of the energy supply after a certain tissue resistance
level is reached.
Tone signal with scope for control of the volume, and an alarm tone in case of faults without facility
for switching off.
Coloured illumination of the LCD display in red, yellow, blue to indicate different operating modes
or error statuses
Facility for tracing output and impedance at the LCD display in the patient circuit on activation
Worldwide application possible through connection scope from 100 to 260 volts, 50 to 60 Hz